GB/T 11606.2-1989 Environmental test methods for analytical instruments Power frequency and voltage test
time:
2024-08-06 11:05:34
- GB/T 11606.2-1989
- Abolished
Standard ID:
GB/T 11606.2-1989
Standard Name:
Environmental test methods for analytical instruments Power frequency and voltage test
Chinese Name:
分析仪器环境试验方法 电源频率与电压试验
Standard category:
National Standard (GB)
-
Date of Release:
1989-03-31 -
Date of Implementation:
1990-03-01 -
Date of Expiration:
2007-12-01
Standard ICS number:
Test >> 19.040 Environmental TestChina Standard Classification Number:
Instruments and meters>>Material composition analysis instruments and environmental monitoring instruments>>N50 Material composition analysis instruments and environmental monitoring instruments comprehensive
alternative situation:
Replaced by GB/T 11606-2007
Review date:
2004-10-14Drafting Organization:
Beijing Analytical Instruments InstituteFocal point Organization:
National Technical Committee for Industrial Process Measurement and Control StandardizationPublishing Department:
China Machinery Industry FederationCompetent Authority:
China Machinery Industry Federation
Skip to download
Summary:
GB/T 11606.2-1989 Environmental test methods for analytical instruments Power frequency and voltage test GB/T11606.2-1989 Standard download decompression password: www.bzxz.net
Some standard content:
National Standard of the People's Republic of China
Environmental test method for analytical instruments
Power frequency and voltage test
The method of environmental test for analytical instrumentsLine frequency and voltage testSubject content and scope of applicationbzxz.net
This standard specifies the conditions and methods for power frequency and voltage test of analytical instruments (hereinafter referred to as instruments). GB11606.2--89
This standard is applicable to the evaluation of the adaptability of instruments to the operation when the power frequency and voltage change. It is also applicable to instruments with special requirements for power frequency and voltage.
2 Test conditions
2.1 The instrument is powered by the power grid, with a power frequency of 50±0.5Hz (or 60±0.5Hz) and a voltage of 220±22V (the waveform distortion factor is not greater than 0.05).
2.2 The adjustment of voltage and frequency must be continuous and smooth. 3: Test method
3.1 Connect the power cord of the instrument to a power supply with adjustable frequency and voltage. 3.2 The power supply voltage is 220, 198, 220, 242V for at least 15 minutes, and the instrument is tested. 3.3 The power supply frequency is 50, 49.5, 50, 50.5Hz (or 60, 59.5, 60, 60.5Hz) for at least 15 minutes, and the instrument is tested.
Approved by the Ministry of Machinery and Electronics Industry of the People's Republic of China on March 25, 1989 and implemented on March 1, 1990
Tip: This standard content only shows part of the intercepted content of the complete standard. If you need the complete standard, please go to the top to download the complete standard document for free.
Environmental test method for analytical instruments
Power frequency and voltage test
The method of environmental test for analytical instrumentsLine frequency and voltage testSubject content and scope of applicationbzxz.net
This standard specifies the conditions and methods for power frequency and voltage test of analytical instruments (hereinafter referred to as instruments). GB11606.2--89
This standard is applicable to the evaluation of the adaptability of instruments to the operation when the power frequency and voltage change. It is also applicable to instruments with special requirements for power frequency and voltage.
2 Test conditions
2.1 The instrument is powered by the power grid, with a power frequency of 50±0.5Hz (or 60±0.5Hz) and a voltage of 220±22V (the waveform distortion factor is not greater than 0.05).
2.2 The adjustment of voltage and frequency must be continuous and smooth. 3: Test method
3.1 Connect the power cord of the instrument to a power supply with adjustable frequency and voltage. 3.2 The power supply voltage is 220, 198, 220, 242V for at least 15 minutes, and the instrument is tested. 3.3 The power supply frequency is 50, 49.5, 50, 50.5Hz (or 60, 59.5, 60, 60.5Hz) for at least 15 minutes, and the instrument is tested.
Approved by the Ministry of Machinery and Electronics Industry of the People's Republic of China on March 25, 1989 and implemented on March 1, 1990
Tip: This standard content only shows part of the intercepted content of the complete standard. If you need the complete standard, please go to the top to download the complete standard document for free.
- Recommended standards
- JB/T 7898-1999 PTFE filled guide rail soft tape
- JGJ/T 21-1993 V-shaped folded plate roof design and construction specification
- HG/T 2750-1996 Indigo Inspection
- JB/T 5178-2002 Technical requirements for hydrostatic steering systems of wheeled tractors
- GB/T 3759-1983 Nuts for ferrule type pipe fittings
- GB 16910-1997 Basic technical conditions for labor hygiene in small industrial enterprises
- SJ 2594-1985 Method for the analysis of boron and metalic impurities in pure SiC14-Spectro-chemical method
- GB/T 15667-1995 Determination of chloride content in fruits, vegetables and their products
- GB 11762-1989 Rapeseed
- GB 12899-1991 Technical requirements for handheld metal detectors
- HG/T 2049-1991 Glass-lined equipment high neck flange
- MT/T 945-2005 Increased safety low voltage cable junction box for coal mines
- JB/T 9008.6-1999 Technical conditions for acceptance of electrical control equipment for wire rope electric hoists
- GB/T 14598.2-1993 Electrical relays with or without electrical relays
- GB/T 3098.20-2004 Mechanical properties of fasteners Butterfly nuts Guaranteed torque
Please remember: "bzxz.net" is the combination of the first letters of the Chinese pinyin of the four Chinese characters "standard download" and the international top-level domain name ".net". ©2024 Standard download websitewww.bzxz.net Mail:bzxznet@163.com