
Methods of measurement for EL2 concentration in semi-insulation Gallium arsenide by infra-red absorption
time:
2024-07-05 09:01:51
- SJ 3249.4-1989
- Abolished
Standard ID:
SJ 3249.4-1989
Standard Name:
Methods of measurement for EL2 concentration in semi-insulation Gallium arsenide by infra-red absorption
Chinese Name:
半绝缘砷化镓中EL2浓度的红外吸收测试方法
Standard category:
Electronic Industry Standard (SJ)
-
Date of Release:
1989-03-20 -
Date of Implementation:
1989-03-25 -
Date of Expiration:
1998-08-01
China Standard Classification Number:
General>>Standardization Management and General Provisions>>A01 Technical Management
alternative situation:
Replaced by GB/T 17170-97

Skip to download
Summary:
SJ 3249.4-1989 Infrared absorption test method for EL2 concentration in semi-insulating gallium arsenide SJ3249.4-1989 standard download decompression password: www.bzxz.net

Some standard content:
Standard of the Ministry of Machinery and Electronics Industry of the People's Republic of China Infrared absorption test method for EL2 concentration in semi-insulating gallium arsenide Subject content and scope of application
SJ3249.4-89
This standard specifies the infrared absorption measurement principle, instrumentation, measurement steps, result calculation and accuracy of deep donor concentration E12 in semi-insulating gallium arsenide.
This standard is applicable to the determination of E12 concentration in undoped semi-insulating gallium arsenide, and is applicable to sample thickness of 2 to 4 mm. It is not suitable for measuring EL2 concentration in chromium-doped semi-insulating samples. 2 Principle
The EI2 deep electron trap light absorption coefficient in semi-insulating GaAs has a corresponding relationship with its concentration. The absorption coefficient at 1.0972μm is measured and the EL2 concentration is calculated by the empirical calibration formula. 3 Instruments and equipment
Spectrophotometer with a range of 0.8~2.5m or wider. 4 Test steps
4.1 Scan with absorption method for zero line calibration. The instrument should be carefully adjusted so that the fluctuation of zero line absorbance in the range of 0.8~2.5μm is no more than ±0.002.
4.2 Place the double-sided polished sample in the light path and align the light beam to the required measurement position. 4.3 Measure the absorption spectrum of the sample in the range of 0.8~2.5μm to obtain the absorbance A-wavelength curve, as shown in the figure: 5.59
Sample original wavelength 3.86cm
lhaioo
Wavelength (×10)μm
Figure Absorption spectrum of typical semi-insulating gallium arsenide sample Approved by the Ministry of Machinery and Electronics Industry of the People's Republic of China on 1989-03-20 and implemented on 1989-03-25
5 Calculation of results
SJ32484-89
5.1 Based on the sample absorption spectrum recorded by the spectrophotometer. Calculate the EL2 concentration by the following formula: Ngl:=1.25×1016α
Where: N—-EL2 concentration (cm)bzxZ.net
EL2 absorption coefficient (cm-1)
αCalculate by the formula below
Where: d sample thickness, cm
-The absorbance values corresponding to 1.0972μm and 2.0μm in the spectrum, 5.2 Calculation of EL2 concentration of typical semi-insulating sample; 5.2.1 The sample thickness is known to be 0.386cm; (1)
5.2.2 The A-input curve obtained by the experiment is shown in the figure, and A is found. and A are 0.487 and 0.293 respectively, 5.2.3α is calculated by the following formula:
The concentration of sample EL2 is
6Report
The report content is as follows
a. Sample source and number
b. Diagram of measurement location,
ca and EL2 data,
α=0.487-0.293×2.30
=1.16(cm-1)
NL21.25×1016×1.16
=1.45×1016(cm-3)
d. Test date and name of tester.
7Accuracy
The measurement accuracy of this method is 10%.
Additional Notes:
This standard was drafted by the 46th Research Institute of the Ministry of Machinery and Electronics Industry. The main drafters of this standard are: Li Guangping, Xun Xiukun, Wang Qin, Zheng Ju, Ge Ping2
Tip: This standard content only shows part of the intercepted content of the complete standard. If you need the complete standard, please go to the top to download the complete standard document for free.
SJ3249.4-89
This standard specifies the infrared absorption measurement principle, instrumentation, measurement steps, result calculation and accuracy of deep donor concentration E12 in semi-insulating gallium arsenide.
This standard is applicable to the determination of E12 concentration in undoped semi-insulating gallium arsenide, and is applicable to sample thickness of 2 to 4 mm. It is not suitable for measuring EL2 concentration in chromium-doped semi-insulating samples. 2 Principle
The EI2 deep electron trap light absorption coefficient in semi-insulating GaAs has a corresponding relationship with its concentration. The absorption coefficient at 1.0972μm is measured and the EL2 concentration is calculated by the empirical calibration formula. 3 Instruments and equipment
Spectrophotometer with a range of 0.8~2.5m or wider. 4 Test steps
4.1 Scan with absorption method for zero line calibration. The instrument should be carefully adjusted so that the fluctuation of zero line absorbance in the range of 0.8~2.5μm is no more than ±0.002.
4.2 Place the double-sided polished sample in the light path and align the light beam to the required measurement position. 4.3 Measure the absorption spectrum of the sample in the range of 0.8~2.5μm to obtain the absorbance A-wavelength curve, as shown in the figure: 5.59
Sample original wavelength 3.86cm
lhaioo
Wavelength (×10)μm
Figure Absorption spectrum of typical semi-insulating gallium arsenide sample Approved by the Ministry of Machinery and Electronics Industry of the People's Republic of China on 1989-03-20 and implemented on 1989-03-25
5 Calculation of results
SJ32484-89
5.1 Based on the sample absorption spectrum recorded by the spectrophotometer. Calculate the EL2 concentration by the following formula: Ngl:=1.25×1016α
Where: N—-EL2 concentration (cm)bzxZ.net
EL2 absorption coefficient (cm-1)
αCalculate by the formula below
Where: d sample thickness, cm
-The absorbance values corresponding to 1.0972μm and 2.0μm in the spectrum, 5.2 Calculation of EL2 concentration of typical semi-insulating sample; 5.2.1 The sample thickness is known to be 0.386cm; (1)
5.2.2 The A-input curve obtained by the experiment is shown in the figure, and A is found. and A are 0.487 and 0.293 respectively, 5.2.3α is calculated by the following formula:
The concentration of sample EL2 is
6Report
The report content is as follows
a. Sample source and number
b. Diagram of measurement location,
ca and EL2 data,
α=0.487-0.293×2.30
=1.16(cm-1)
NL21.25×1016×1.16
=1.45×1016(cm-3)
d. Test date and name of tester.
7Accuracy
The measurement accuracy of this method is 10%.
Additional Notes:
This standard was drafted by the 46th Research Institute of the Ministry of Machinery and Electronics Industry. The main drafters of this standard are: Li Guangping, Xun Xiukun, Wang Qin, Zheng Ju, Ge Ping2
Tip: This standard content only shows part of the intercepted content of the complete standard. If you need the complete standard, please go to the top to download the complete standard document for free.
- Recommended standards
- GA/T 440-2003 Interface between vehicle anti-theft alarm equipment and vehicle wireless communication terminal equipment in vehicle anti-robbery and anti-theft network alarm system
- GB/T 3769-1983 Scales and sizes for plotting frequency characteristic and polar diagrams
- JB/T 6923-1993 Vacuum pressure impregnation equipment
- GB 17511.1-1998 Food additive Allura red
- YY/T 0456.2-2003 Reagents for blood cell analyzers Part 2: Hemolytic agents
- GB/T 24041-2000 Determination of the purpose and scope of environmental management life cycle assessment and inventory analysis
- JB/T 5992.1-1992 General principles for classification and coding of mechanical manufacturing process methods
- JB/T 7556-1994 Technical requirements for 2 1/4Cr-1Mo steel forgings for hot wall hydrogenation reactors
- JB/T 8866-2001 Cartridge pressure filter
- GB/T 12533-1990 Determinaiton of micellaneous matter in edible fungi
- GA/T 388-2002 Technical requirements for computer information system security level protection operating system
- GB/T 19476-2004 Hard-copy output of engineering drawings—Specification for the structure of control files
- GB/T 4700.5-1998 Chemical analysis methods for silicon-calcium alloys - Determination of carbon content by infrared absorption method
- GB/T 9852.2-1988 Standards for the archives of the national multipurpose investigation of the coastal zone and tidal wetland resources-Regulations on bibliographical description
- JB/T 9845-1999 Rules for compiling instrument models for tractors and internal combustion engines
Please remember: "bzxz.net" is the combination of the first letters of the Chinese pinyin of the four Chinese characters "standard download" and the international top-level domain name ".net". ©2024 Standard download websitewww.bzxz.net Mail:bzxznet@163.com