GB 5579-1985 High-voltage cable plug and socket connections for medical x-ray equipment
time:
1985-10-31 16:00:00
- GB 5579-1985
- in force
Standard ID:
GB 5579-1985
Standard Name:
High-voltage cable plug and socket connections for medical x-ray equipment
Chinese Name:
医用x射线设备高压电缆插头插座连接
Standard category:
National Standard (GB)
-
Date of Release:
1985-10-31 -
Date of Implementation:
1986-07-01
Standard ICS number:
Medical and Health Technology>>Medical Equipment>>11.040.50 Radiography EquipmentChina Standard Classification Number:
Electrician>>Electrical Materials and General Parts>>K13 Cable and Accessories
alternative situation:
Replaced WS 2-166-1980; replaced by GB/T 10151-2008Procurement status:
IEC 526-78
publishing house:
China Standard PressISBN:
155066.1-3877Publication date:
1986-07-01
Release date:
1985-10-31Review date:
2004-10-14Drafting Organization:
southwest medical equipment factoryFocal point Organization:
National Technical Committee for Standardization of Medical Electrical AppliancesPublishing Department:
National Bureau of StandardsCompetent Authority:
State Food and Drug Administration

Skip to download
Summary:
This standard applies to three-core and four-core high-voltage cable plugs and sockets for medical X-ray equipment. This component is a high-voltage connector for medical X-ray equipment. This standard is equivalent to the international standard IEC 526 (1978) "Connection of high-voltage cable plugs and sockets for medical X-ray equipment". GB 5579-1985 High-voltage cable plug and socket connection for medical x-ray equipment GB5579-1985 Standard download and decompression password: www.bzxz.net

Some standard content:
National Standards of the People's Republic of China
High-voltage cable plug and socket connections for medical X-ray equipment
UDC621.316.541
:621.3.027.31| |tt||.7:616-7
GB5579-85
This standard applies to three-core and four-core high-voltage cable plugs and sockets for medical X-ray equipment. This component is for medical X-ray equipment. High voltage connector.
This standard is equivalent to the international standard IEC526 (1978) "Connection of high-voltage cable plugs and sockets for medical X-ray equipment". Type and basic dimensions
1.1 High-voltage cable plug and socket are composed of the following components: 1.1.1 High-voltage cable plug assembly is composed of the following main parts: a. Ring nut;
b. Plug body;
Install the above parts at the end of the prepared cable to ground, insulate and seal them to form the plug assembly. 1.1.2 The jack socket assembly consists of the following main parts: a: mounting flange;
b. thrust ring
socket with jack terminals.
The mounting flange may be an integral part of the high-voltage generator, X-ray tube assembly, or other part of the equipment. 1.2 The basic dimensions and types of high-voltage cable plugs and sockets should comply with the requirements in Figures 1 to 6. Released by the National Bureau of Standards on 1985-10-31
Implemented on 1986-07-01
1||tt| |R4
GB5579—85
417.4.±0.05
$71.5±0.6
55 ±2
R(1.6±0.4)
Φ3.96±0.02||tt ||33.4±0.2
(9001)
ISO263
2 yuan-20UN
1--ring nut; 2 one plug Body; 3 pins Figure 1 Three-core cable end plug
Pin markings specified in Table 1
(291)
(*0-s°9)
(R1.6±0.4)
(s°9)
( 1+8)
GB5579-85
(@61.5max)C
$35.7±0.2
(49.3max)||tt| |$34.1±0.2
Φ17.4±0.05
37.3±0.3
ISO262 | tt||(20)
Figure 2 three-hole socket with jack binding posts
GB5579—85
No. 2-20UN
1--Thrust ring, 2-mounting flange
Picture 3 Jack socket assembly
2 inch-20UN
(#52.4± 0.6)
$44.5±0.4
Figure 4 thrust ring
(120°)
ISO263
(1 +9-9)
GB5579--85
36.8±0.1
$71.5± 0.9
(R2±0.5)
455±2
No. 2-20UN
R(1.6±0.4)|| tt||R1.5*g.5
$3.96±0.02
$33.4±0.2
(90+0T)
1SO263|| tt||1-ring nut, 2-plug body, 3-pin Figure 5 Four-core cable end plug
Select the pin mark according to Table 2
xe9||tt| |()
(59s*9)
(R1.6±0.4)
( s°9)
GB5579—85
$35.7±0.2|| tt||(@49.3max
634.1±0.
ISO262
37.3±0.3
(58.5-0.3)||tt| |Figure 6 Four-hole socket with jack binding posts
A direction
Socket binding post markings specified in Table 2
2 mark
GB5579—85||tt ||2.1 The markings of the pins and socket terminals of the three-core cable plug should use graphic symbols or letter symbols. The graphic symbols and letter symbols must comply with the symbols and marking positions shown in Table 1 and Figure 1 and Figure 2. Table 1||tt| |Three-core cable plug pins and
socket binding post mark
graphic symbol
letter symbol
public
none|| tt||C
Large focus filament
Small focus filament
2.2 four-core cable plug pins and The signs of socket terminals should use graphic symbols or letter symbols. The graphic symbols and letter symbols must comply with the symbols and marking positions shown in Table 2 and Figures 5 and 6. Table 2
four-core cable plug pins and
socket binding post mark
graphic symbol
letter symbol
2.3 Factory name mark
Public
Large focus filament
Each set of plug and socket components should be marked with the factory name mark at the appropriate location. 3 connections
small focal length filament
3.1. The wire connection between the high-voltage generator and the socket terminal must be connected as shown in Table 3. Auxiliary function potential
Socket
Terminal
Common end
Large focal length filament
L| |tt||Small focus filament
Auxiliary potential
Connection diagram
Common C
Large focus L|| tt||Xiaojiao S
GB5579-85
Table 3
Wire connection between high voltage generator and three-hole socket binding post
Cathode end with filament power supply|| |tt||Anode end
Bifocal or single focus tube
Dual filament L and S
One end of the power supply
Filament L power supply
One end
The other end of the filament S power supply
The other end
Single focus tube
One end of the filament power supply
The other end of the filament power supply| | tt | Wire connection for socket terminals
with filament power supply and auxiliary potential
cathode end
bifocal or single focus tube
double filament L, S||tt| |Power supply and auxiliary power
One end of bit G
The other end of filament L power supply
The other end of filament S power supply
The other end||tt| |The
other end of the auxiliary potential G
single focus tube
end of the filament power supply and the one end of the auxiliary potential
filament L, S short bZxz.net
interconnection filament electrical
source
auxiliary potential G
other end
3.2 X-ray tube assembly with The wire connections of the jack and socket binding posts must be connected as shown in Table 4. *If necessary, make auxiliary functional potential.
Anode terminal
All socket terminals
Partial short circuit
Socket
Common terminal
c| |tt||Large focus filament
Small focus filament
Auxiliary potential
Device G
Connection diagram|| tt||Public C
Large Jiao L
Small Jiao S
GB5579-85
Table 4
X-ray tube assembly and three-hole socket wiring Column
wire connection
cathode end with filament
anode end
double focus tube
double filament L, S
One end of
The other end of filament L
The other end of filament S
Single focus tube
X-ray tube assembly and four-hole socket binding post| |tt||Wire connection
Cathode end with filament and auxiliary device
Anode end
Double focus tube
Double filament L, S
One end of the filament
The other end of the filament
The other end of the filament or
all Socket wiring
one end of the post connection
end of the auxiliary device
the other end of the filament L
The other
end of the filament S
the other
one end of the auxiliary device
3.3 The connection method of the cable and the plug assembly at both ends must be connected as shown in Table 5. *If necessary, make auxiliary functional potential.
Single focus tube
One end of the filament
One end of the auxiliary device
The other end of the filament
The other end of the filament| |tt||Other side of auxiliary device
One end
All socket wiring
Pole connection
Tip: This standard content only shows part of the intercepted content of the complete standard. If you need the complete standard, please go to the top to download the complete standard document for free.
High-voltage cable plug and socket connections for medical X-ray equipment
UDC621.316.541
:621.3.027.31| |tt||.7:616-7
GB5579-85
This standard applies to three-core and four-core high-voltage cable plugs and sockets for medical X-ray equipment. This component is for medical X-ray equipment. High voltage connector.
This standard is equivalent to the international standard IEC526 (1978) "Connection of high-voltage cable plugs and sockets for medical X-ray equipment". Type and basic dimensions
1.1 High-voltage cable plug and socket are composed of the following components: 1.1.1 High-voltage cable plug assembly is composed of the following main parts: a. Ring nut;
b. Plug body;
Install the above parts at the end of the prepared cable to ground, insulate and seal them to form the plug assembly. 1.1.2 The jack socket assembly consists of the following main parts: a: mounting flange;
b. thrust ring
socket with jack terminals.
The mounting flange may be an integral part of the high-voltage generator, X-ray tube assembly, or other part of the equipment. 1.2 The basic dimensions and types of high-voltage cable plugs and sockets should comply with the requirements in Figures 1 to 6. Released by the National Bureau of Standards on 1985-10-31
Implemented on 1986-07-01
1||tt| |R4
GB5579—85
417.4.±0.05
$71.5±0.6
55 ±2
R(1.6±0.4)
Φ3.96±0.02||tt ||33.4±0.2
(9001)
ISO263
2 yuan-20UN
1--ring nut; 2 one plug Body; 3 pins Figure 1 Three-core cable end plug
Pin markings specified in Table 1
(291)
(*0-s°9)
(R1.6±0.4)
(s°9)
( 1+8)
GB5579-85
(@61.5max)C
$35.7±0.2
(49.3max)||tt| |$34.1±0.2
Φ17.4±0.05
37.3±0.3
ISO262 | tt||(20)
Figure 2 three-hole socket with jack binding posts
GB5579—85
No. 2-20UN
1--Thrust ring, 2-mounting flange
Picture 3 Jack socket assembly
2 inch-20UN
(#52.4± 0.6)
$44.5±0.4
Figure 4 thrust ring
(120°)
ISO263
(1 +9-9)
GB5579--85
36.8±0.1
$71.5± 0.9
(R2±0.5)
455±2
No. 2-20UN
R(1.6±0.4)|| tt||R1.5*g.5
$3.96±0.02
$33.4±0.2
(90+0T)
1SO263|| tt||1-ring nut, 2-plug body, 3-pin Figure 5 Four-core cable end plug
Select the pin mark according to Table 2
xe9||tt| |()
(59s*9)
(R1.6±0.4)
( s°9)
GB5579—85
$35.7±0.2|| tt||(@49.3max
634.1±0.
ISO262
37.3±0.3
(58.5-0.3)||tt| |Figure 6 Four-hole socket with jack binding posts
A direction
Socket binding post markings specified in Table 2
2 mark
GB5579—85||tt ||2.1 The markings of the pins and socket terminals of the three-core cable plug should use graphic symbols or letter symbols. The graphic symbols and letter symbols must comply with the symbols and marking positions shown in Table 1 and Figure 1 and Figure 2. Table 1||tt| |Three-core cable plug pins and
socket binding post mark
graphic symbol
letter symbol
public
none|| tt||C
Large focus filament
Small focus filament
2.2 four-core cable plug pins and The signs of socket terminals should use graphic symbols or letter symbols. The graphic symbols and letter symbols must comply with the symbols and marking positions shown in Table 2 and Figures 5 and 6. Table 2
four-core cable plug pins and
socket binding post mark
graphic symbol
letter symbol
2.3 Factory name mark
Public
Large focus filament
Each set of plug and socket components should be marked with the factory name mark at the appropriate location. 3 connections
small focal length filament
3.1. The wire connection between the high-voltage generator and the socket terminal must be connected as shown in Table 3. Auxiliary function potential
Socket
Terminal
Common end
Large focal length filament
L| |tt||Small focus filament
Auxiliary potential
Connection diagram
Common C
Large focus L|| tt||Xiaojiao S
GB5579-85
Table 3
Wire connection between high voltage generator and three-hole socket binding post
Cathode end with filament power supply|| |tt||Anode end
Bifocal or single focus tube
Dual filament L and S
One end of the power supply
Filament L power supply
One end
The other end of the filament S power supply
The other end
Single focus tube
One end of the filament power supply
The other end of the filament power supply| | tt | Wire connection for socket terminals
with filament power supply and auxiliary potential
cathode end
bifocal or single focus tube
double filament L, S||tt| |Power supply and auxiliary power
One end of bit G
The other end of filament L power supply
The other end of filament S power supply
The other end||tt| |The
other end of the auxiliary potential G
single focus tube
end of the filament power supply and the one end of the auxiliary potential
filament L, S short bZxz.net
interconnection filament electrical
source
auxiliary potential G
other end
3.2 X-ray tube assembly with The wire connections of the jack and socket binding posts must be connected as shown in Table 4. *If necessary, make auxiliary functional potential.
Anode terminal
All socket terminals
Partial short circuit
Socket
Common terminal
c| |tt||Large focus filament
Small focus filament
Auxiliary potential
Device G
Connection diagram|| tt||Public C
Large Jiao L
Small Jiao S
GB5579-85
Table 4
X-ray tube assembly and three-hole socket wiring Column
wire connection
cathode end with filament
anode end
double focus tube
double filament L, S
One end of
The other end of filament L
The other end of filament S
Single focus tube
X-ray tube assembly and four-hole socket binding post| |tt||Wire connection
Cathode end with filament and auxiliary device
Anode end
Double focus tube
Double filament L, S
One end of the filament
The other end of the filament
The other end of the filament or
all Socket wiring
one end of the post connection
end of the auxiliary device
the other end of the filament L
The other
end of the filament S
the other
one end of the auxiliary device
3.3 The connection method of the cable and the plug assembly at both ends must be connected as shown in Table 5. *If necessary, make auxiliary functional potential.
Single focus tube
One end of the filament
One end of the auxiliary device
The other end of the filament
The other end of the filament| |tt||Other side of auxiliary device
One end
All socket wiring
Pole connection
Tip: This standard content only shows part of the intercepted content of the complete standard. If you need the complete standard, please go to the top to download the complete standard document for free.
- Recommended standards
- GB/T 40036-2021 Translation services—Post-editing of machine translation output—Requirements
- JB/T 7763.1-1995 Technical requirements for starting fluids and devices for internal combustion engines
- GB 30000.24-2013 Rules for classification and labelling of chemicals—Part 24:Reproductive toxicity
- QB/T 2254-1996 Sewing machine thread
- NY/T 646-2002 Specification for quality evaluation of screw conveyors
- GB/T 9074.28-1988 Conical locking washers for assemblies
- GB 16541-1996 Safety technical requirements for shaft cage hoisting signal systems
- JB/T 9161.10-1999 Dimensions of screws for internal grinding machines
- SY/T 5379-1991 Technical requirements for drilling fluid shear force gauges
- CJJ 26-1989 Technical Grade Standard for House Construction Workers CJJ26-89
- HG/T 3135-1998 Technical requirements for fully automatic sodium ion exchangers
- GB 8241-1987 Terminology of ship cabin equipment
- JB/T 8710-1998 Hydraulic explosion-proof hoist and hoist winch
- GB/T 2423.47-1997 Environmental testing for electric and electronic products Part 2: Test methods Test Fg: Acoustic vibration
- GB/T 4958.17-1992 Measurement methods for equipment used in terrestrial radio-relay systems Part 2: Subsystem measurements Section 6: Diversity, two-way and hot standby equipment
Please remember: "bzxz.net" is the combination of the first letters of the Chinese pinyin of the four Chinese characters "standard download" and the international top-level domain name ".net". ©2024 Standard download websitewww.bzxz.net Mail:[email protected]