GB/T 11606.16-1989 Environmental test methods for analytical instruments - Drop test
time:
2024-08-06 10:57:13
- GB/T 11606.16-1989
- Abolished
Standard ID:
GB/T 11606.16-1989
Standard Name:
Environmental test methods for analytical instruments - Drop test
Chinese Name:
分析仪器环境试验方法 跌落试验
Standard category:
National Standard (GB)
-
Date of Release:
1989-03-31 -
Date of Implementation:
1990-03-01 -
Date of Expiration:
2007-12-01
Standard ICS number:
Test >> 19.040 Environmental TestChina Standard Classification Number:
Instruments and meters>>Material composition analysis instruments and environmental monitoring instruments>>N50 Material composition analysis instruments and environmental monitoring instruments comprehensive
alternative situation:
Replaced by GB/T 11606-2007
Review date:
2004-10-14Drafting Organization:
Beijing Analytical Instruments InstituteFocal point Organization:
National Technical Committee for Industrial Process Measurement and Control StandardizationPublishing Department:
China Machinery Industry FederationCompetent Authority:
China Machinery Industry Federation
Skip to download
Summary:
GB/T 11606.16-1989 Environmental test methods for analytical instruments Drop test GB/T11606.16-1989 Standard download decompression password: www.bzxz.net
Some standard content:
National Standard of the People's Republic of China
Environmental test method for analytical instruments
Drop test
The method of environmental test for analytical instrumentsFall duwn test
1 Subject content and scope of application
This standard specifies the drop test conditions and test methods for analytical instruments (hereinafter referred to as instruments). This standard is applicable to the assessment of the protection capability of transport packaging for instruments. 2 Test conditions
The height of the drop is specified according to the gross weight of the instrument after packaging, see the following table: Package weight
>100~200
3 Test method
3.1 Free drop test
3.1.1 Test table
Flat and hard cement floor or steel plate table, 3.1.2 Drop method
The bottom of the package is horizontal and falls in a free fall manner. 3.1.3 Number of drops
3.2 Inclined drop
3.2.1 Test table
Flat and hard cement floor or steel table. Drop method
Free drop
Tilt drop
GB11606.16--89
Drop height
250mm or 30
3.2.2 Test conditions
When the length of the edge of the bottom surface of the package is less than 500mm, one edge is in contact with the ground and the bottom surface inclination angle is 30. When the length of the bottom surface shuttle edge is greater than or equal to 500mm, one edge is in contact with the ground and the distance between the inclined bottom and the ground is 250mm. 3.2.3 Drop method
Each edge of the bottom surface of the package is tested once according to the provisions of Article 3.2.2. Final test
Inspect the instrument according to the relevant standards. Approved by the Ministry of Machinery and Electronics Industry of the People's Republic of China on March 25, 198932
Implemented on March 1, 1990bzxz.net
GB/T11606.16-1989 "Analytical Instrument Environmental Test Method Drop Test" No. 1 Amendment
This amendment was approved and issued by the State Administration of Technical Supervision on July 19, 2000, and implemented on December 1, 2000. Published in the 9th issue of China Standardization in 2000). 1. In Chapter 2 (including tables), "drop height" is changed to "drop conditions". In the table, "tip-over drop" is changed to "inclined drop". 2. Supplement: 3.1 Initial inspection
Inspect the instrument according to the relevant standards. 3.3.2 "inclined drop\" is changed to inclined drop test. 4. The order of the original 3.1 and the following articles is modified accordingly. The punctuation mark in 5.3.2.2 should be changed to "one edge is touching the ground, and the bottom inclination angle is 30;\.
Tip: This standard content only shows part of the intercepted content of the complete standard. If you need the complete standard, please go to the top to download the complete standard document for free.
Environmental test method for analytical instruments
Drop test
The method of environmental test for analytical instrumentsFall duwn test
1 Subject content and scope of application
This standard specifies the drop test conditions and test methods for analytical instruments (hereinafter referred to as instruments). This standard is applicable to the assessment of the protection capability of transport packaging for instruments. 2 Test conditions
The height of the drop is specified according to the gross weight of the instrument after packaging, see the following table: Package weight
>100~200
3 Test method
3.1 Free drop test
3.1.1 Test table
Flat and hard cement floor or steel plate table, 3.1.2 Drop method
The bottom of the package is horizontal and falls in a free fall manner. 3.1.3 Number of drops
3.2 Inclined drop
3.2.1 Test table
Flat and hard cement floor or steel table. Drop method
Free drop
Tilt drop
GB11606.16--89
Drop height
250mm or 30
3.2.2 Test conditions
When the length of the edge of the bottom surface of the package is less than 500mm, one edge is in contact with the ground and the bottom surface inclination angle is 30. When the length of the bottom surface shuttle edge is greater than or equal to 500mm, one edge is in contact with the ground and the distance between the inclined bottom and the ground is 250mm. 3.2.3 Drop method
Each edge of the bottom surface of the package is tested once according to the provisions of Article 3.2.2. Final test
Inspect the instrument according to the relevant standards. Approved by the Ministry of Machinery and Electronics Industry of the People's Republic of China on March 25, 198932
Implemented on March 1, 1990bzxz.net
GB/T11606.16-1989 "Analytical Instrument Environmental Test Method Drop Test" No. 1 Amendment
This amendment was approved and issued by the State Administration of Technical Supervision on July 19, 2000, and implemented on December 1, 2000. Published in the 9th issue of China Standardization in 2000). 1. In Chapter 2 (including tables), "drop height" is changed to "drop conditions". In the table, "tip-over drop" is changed to "inclined drop". 2. Supplement: 3.1 Initial inspection
Inspect the instrument according to the relevant standards. 3.3.2 "inclined drop\" is changed to inclined drop test. 4. The order of the original 3.1 and the following articles is modified accordingly. The punctuation mark in 5.3.2.2 should be changed to "one edge is touching the ground, and the bottom inclination angle is 30;\.
Tip: This standard content only shows part of the intercepted content of the complete standard. If you need the complete standard, please go to the top to download the complete standard document for free.
- Recommended standards
- JB/T 3585.2-1996 Induction synchronizer types and basic dimensions
- SB/T 10335-2000 Outlet management specifications general requirement for supermarket and convenience store company management
- GB/T 38171-2019 Metal chelating chromatographic medium
- GB/T 5009.51-2003 Analytical methods for hygienic standards of non-fermented soy products and gluten
- GB/T 9212-1988 Carbide torch-shaped rotary files
- QB/T 2419-1998 Roller grass cutter
- GB/T 15555.8-1995 Determination of total chromium in solid wastes - Ammonium ferrous sulfate titration method
- GB/T 8762.8-2000 Chemical analysis method for europium oxide - Determination of lanthanum oxide, cerium oxide, praseodymium oxide, neodymium oxide, samarium oxide, gadolinium oxide, terbium oxide, dysprosium oxide, holmium oxide, erbium oxide, thulium oxide, ytterbium oxide, lutetium oxide and tantalum oxide in europium oxide by inductively coupled plasma atomic emission spectrometry
- GB/T 6882-2008 Acoustics - Determination of sound power levels of noise sources using pressure - Precision methods for anechoic and hemi-anechoic rooms
- MT/T 440-1995 Method for determination of mine ventilation resistance
- JB/T 6621-1993 Determination method of linear expansion coefficient of flexible graphite plate
- CECS : 21-2000 Technical specification for ultrasonic detection of concrete defects CECS21:2000
- JB/T 8487-1996 Piston lathe accuracy inspection
- GB 2689.4-1981 Best linear unbiased estimation method for life test and accelerated life test (for Weibull distribution)
- GB/T 9032-1988 Technical requirements for pulse push-button telephones