GB/T 5177.3-1985 Determination of average relative molecular weight of industrial alkylbenzenes - Gas-liquid chromatography
time:
2024-08-04 20:44:42
- GB/T 5177.3-1985
- Abolished
Standard ID:
GB/T 5177.3-1985
Standard Name:
Determination of average relative molecular weight of industrial alkylbenzenes - Gas-liquid chromatography
Chinese Name:
工业烷基苯平均相对分子量的测定 气液色谱法
Standard category:
National Standard (GB)
-
Date of Release:
1985-05-10 -
Date of Implementation:
1986-01-01 -
Date of Expiration:
2008-12-01
Standard ICS number:
Chemical Technology>>Organic Chemistry>>71.080.15 Aromatic HydrocarbonsChina Standard Classification Number:
Chemicals>>Organic Chemical Raw Materials>>G17 General Organic Chemical Raw Materials
alternative situation:
Replaced by GB/T 5177-2008
Release date:
1985-05-10Review date:
2004-10-14Drafting Organization:
Daily Chemicals Institute of the Ministry of Light IndustryFocal point Organization:
National Technical Committee on Standardization of Surfactants and DetergentsPublishing Department:
State Administration of StandardsCompetent Authority:
China Light Industry Federation

Skip to download
Summary:
This standard is applicable to the determination of the average relative molecular weight of industrial linear alkylbenzenes, but not to branched alkylbenzenes. GB/T 5177.3-1985 Determination of the average relative molecular weight of industrial alkylbenzenes - Gas-liquid chromatography method GB/T5177.3-1985 Standard download decompression password: www.bzxz.net

Some standard content:
National Standard of the People's Republic of China
Determitation of mean relative molecular mass forindustrial alkylbenzene-Gas-liquld chromatographic methodUDC 547.52: 541
GB 5177.3—85
This standard is applicable to the determination of mean relative molecular mass for industrial straight-chain alkylbenzenes, but not for branched-chain alkylbenzenes. 1 Principle
Gas-liquid chromatography is used to determine the carbon number distribution of industrial straight-chain alkylbenzenes and calculate the average relative molecular mass. 2 Instruments
2.1 Chromatograph
2.1.1 Filling column: stainless steel column, 1-6 m long, 2-4 mm inner diameter, filled with high temperature non-polar stationary phase (such as SE300V!01), pre-aged before the test, or glass capillary tube, 20-50 m long. The chromatographic peaks of alkyl fluorine with different carbon atoms should be well separated.
2.1.2 Flame ionization detector.
2.1.3 Electronic integrator or integrator. If these two instruments are not available, the paper cutting weighing method can be used. 2, 1.4 Recorder.
2.1.5 Carrier gas: nitrogen.
2.2 Micro syringe: 1 or 10 μl.
3 Reagents and reference samples
3.1 Reference alkylbenzene: a mixture of alkylbenzenes with known chain lengths, for example, C15 to C15 alkylbenzenes. A product with known composition can be used as a reference sample. This reference sample can be used to verify whether the performance of the chromatograph is normal. 3.2 Acetone: analytical grade.
4 Chromatographic analysis conditions
4.1 Column temperature
4.1.1 Column temperature: based on the properties of the sample, the temperature is between 150 and 180°C. 4.1.2 Program heating: the initial temperature is 140 to 170°C, the heating rate is 2 to 2 minutes, and the final temperature is 180 to 250°C. 4.2 Injection port temperature, 250 to 300°C. 4.3 Carrier gas: depending on the rod type and diameter, the flow rate can be between 0.3 and 40 m/min. 5 Chromatographic analysis
5.1 Preparation of sample
If a capillary column (2.1.1) is used, a mixture of 1 volume of alkyl amine and 1 volume of internal amine (3.2) can be prepared if necessary. Alkyl amine can be used directly when using a packed column (2.1.1). National Bureau of Standards Issued on May 10, 1985
Implemented on January 1, 1986
5. 2 Introduction of samplewww.bzxz.net
GB PTA-BS
Use a syringe (2.2) to inject a sufficient amount of sample (5.1) into the chromatograph so that the obtained chromatogram has an appropriate height. Typical chromatograms are shown in Figures 1 and 2.
市-Cm
.Fig.1 Column chromatogram of alkylbenzene (the peak top number is the substitution position of benzene on the alkyl chain, representing phenyl)6,5
Fig.2 Capillary column chromatogram of alkyl (the middle represents the alkyl)6 Qualitative and quantitative methods
6.1 Qualitative
GB 5177.3-85
The components of the sample are determined by comparing the sample chromatogram with the chromatogram of alkylbenzene (3.1). 6.2 Quantitative
Use an electronic integrator or a quadrature meter (2.1.3) or a paper-cut weighing method to determine the peak area of alkylbenzenes with different carbon atoms, and calculate the total peak area. Expression of results
7.1 Calculation method
The percentage of the peak area of alkylbenzene with i carbon atoms to the total area is calculated by formula (【): B,=gix100
The average relative molecular weight of alkylbenzene is calculated by formula (2): M=_100
Wherein: B. The percentage of the peak area of alkylbenzene with i carbon atoms to the total peak area; The average relative molecular weight of alkylbenzene is:
peak area of alkylbenzene with i carbon atoms,
total peak area,
molecular weight of alkylbenzene with i carbon atoms, see the table below.
7.2 Precision
The same alkylbenzene sample was compared and analyzed in 10 laboratories, and the following statistical results were obtained: Average value (average relative molecular weight) 242.6 Repeatability standard deviation
Reproducibility standard deviation
Additional instructions:
This standard was proposed by the Ministry of Light Industry of the People's Republic of China and is technically managed by the Daily Chemical Industry Science Research Institute of the Ministry of Light Industry. This standard was drafted by the Daily Chemical Industry Science Research Institute of the Ministry of Light Industry. The main drafter of this standard is Peng Xianqun.
Tip: This standard content only shows part of the intercepted content of the complete standard. If you need the complete standard, please go to the top to download the complete standard document for free.
Determitation of mean relative molecular mass forindustrial alkylbenzene-Gas-liquld chromatographic methodUDC 547.52: 541
GB 5177.3—85
This standard is applicable to the determination of mean relative molecular mass for industrial straight-chain alkylbenzenes, but not for branched-chain alkylbenzenes. 1 Principle
Gas-liquid chromatography is used to determine the carbon number distribution of industrial straight-chain alkylbenzenes and calculate the average relative molecular mass. 2 Instruments
2.1 Chromatograph
2.1.1 Filling column: stainless steel column, 1-6 m long, 2-4 mm inner diameter, filled with high temperature non-polar stationary phase (such as SE300V!01), pre-aged before the test, or glass capillary tube, 20-50 m long. The chromatographic peaks of alkyl fluorine with different carbon atoms should be well separated.
2.1.2 Flame ionization detector.
2.1.3 Electronic integrator or integrator. If these two instruments are not available, the paper cutting weighing method can be used. 2, 1.4 Recorder.
2.1.5 Carrier gas: nitrogen.
2.2 Micro syringe: 1 or 10 μl.
3 Reagents and reference samples
3.1 Reference alkylbenzene: a mixture of alkylbenzenes with known chain lengths, for example, C15 to C15 alkylbenzenes. A product with known composition can be used as a reference sample. This reference sample can be used to verify whether the performance of the chromatograph is normal. 3.2 Acetone: analytical grade.
4 Chromatographic analysis conditions
4.1 Column temperature
4.1.1 Column temperature: based on the properties of the sample, the temperature is between 150 and 180°C. 4.1.2 Program heating: the initial temperature is 140 to 170°C, the heating rate is 2 to 2 minutes, and the final temperature is 180 to 250°C. 4.2 Injection port temperature, 250 to 300°C. 4.3 Carrier gas: depending on the rod type and diameter, the flow rate can be between 0.3 and 40 m/min. 5 Chromatographic analysis
5.1 Preparation of sample
If a capillary column (2.1.1) is used, a mixture of 1 volume of alkyl amine and 1 volume of internal amine (3.2) can be prepared if necessary. Alkyl amine can be used directly when using a packed column (2.1.1). National Bureau of Standards Issued on May 10, 1985
Implemented on January 1, 1986
5. 2 Introduction of samplewww.bzxz.net
GB PTA-BS
Use a syringe (2.2) to inject a sufficient amount of sample (5.1) into the chromatograph so that the obtained chromatogram has an appropriate height. Typical chromatograms are shown in Figures 1 and 2.
市-Cm
.Fig.1 Column chromatogram of alkylbenzene (the peak top number is the substitution position of benzene on the alkyl chain, representing phenyl)6,5
Fig.2 Capillary column chromatogram of alkyl (the middle represents the alkyl)6 Qualitative and quantitative methods
6.1 Qualitative
GB 5177.3-85
The components of the sample are determined by comparing the sample chromatogram with the chromatogram of alkylbenzene (3.1). 6.2 Quantitative
Use an electronic integrator or a quadrature meter (2.1.3) or a paper-cut weighing method to determine the peak area of alkylbenzenes with different carbon atoms, and calculate the total peak area. Expression of results
7.1 Calculation method
The percentage of the peak area of alkylbenzene with i carbon atoms to the total area is calculated by formula (【): B,=gix100
The average relative molecular weight of alkylbenzene is calculated by formula (2): M=_100
Wherein: B. The percentage of the peak area of alkylbenzene with i carbon atoms to the total peak area; The average relative molecular weight of alkylbenzene is:
peak area of alkylbenzene with i carbon atoms,
total peak area,
molecular weight of alkylbenzene with i carbon atoms, see the table below.
7.2 Precision
The same alkylbenzene sample was compared and analyzed in 10 laboratories, and the following statistical results were obtained: Average value (average relative molecular weight) 242.6 Repeatability standard deviation
Reproducibility standard deviation
Additional instructions:
This standard was proposed by the Ministry of Light Industry of the People's Republic of China and is technically managed by the Daily Chemical Industry Science Research Institute of the Ministry of Light Industry. This standard was drafted by the Daily Chemical Industry Science Research Institute of the Ministry of Light Industry. The main drafter of this standard is Peng Xianqun.
Tip: This standard content only shows part of the intercepted content of the complete standard. If you need the complete standard, please go to the top to download the complete standard document for free.
- Recommended standards
- SN 0161-1992 Method for determination of parathion methyl residue in fruit for export
- SJ/T 10624-1995 Methods of life test for X-ray tubes
- DA/T 14-1994 Specification for the compilation of the guide to archives
- HG/T 2049-1991 Glass-lined equipment high neck flange
- GB/T 15131.3-1995 Information processing data exchange 130 mm modified frequency modulation recording disks with a bit density of 13262 flux reversals/radian and 80 tracks per side Part 3: Track format B (for 80 tracks)
- GB 17052-1997 Hygienic standard for rock wool dust in workshop air
- GB 4797.3-1986 Natural environmental conditions for electrical and electronic products
- JB/T 8739-1998 Flameproof high voltage distribution device for mines
- JB/T 8486.2-1996 Technical requirements for CNC universal tool milling machines
- JB/T 8824.3-1998 Technical requirements for unified thread taps
- GB 2508-1989 Marine lap-welded loose-fitting steel flange (four-position)
- CB/T 3559-1994 Manual ultrasonic flaw detection process and quality classification of ship steel welds
- GB 8951-1988 Hygienic Standards for Liquor Factory
- GB/T 3939.1-2004 Naming and marking of main fishing gear materials - Fishing line
- GB/T 3900-2003 Tire valve series
Please remember: "bzxz.net" is the combination of the first letters of the Chinese pinyin of the four Chinese characters "standard download" and the international top-level domain name ".net". ©2024 Standard download websitewww.bzxz.net Mail:bzxznet@163.com