GB/T 3389.3-2001 Test methods for properties of piezoelectric ceramic materials - Test of Curie temperature Tc
time:
2024-08-10 23:34:56
- GB/T 3389.3-2001
- Abolished
Standard ID:
GB/T 3389.3-2001
Standard Name:
Test methods for properties of piezoelectric ceramic materials - Test of Curie temperature Tc
Chinese Name:
压电陶瓷材料性能试验方法 居里温度Tc的测试
Standard category:
National Standard (GB)
-
Date of Release:
2001-11-16 -
Date of Implementation:
2002-08-01 -
Date of Expiration:
2009-02-01
Standard ICS number:
31.030China Standard Classification Number:
Electronic Components and Information Technology>>Special Materials, Parts, and Structural Components for Electronic Equipment>>Special Materials for L90 Electronic Technology
alternative situation:
Replaced GB/T 3389.3-1982; replaced by GB/T 3389-2008
publishing house:
China Standards PressISBN:
155066.1-18283Publication date:
2002-08-01
Release date:
1982-12-30Review date:
2004-10-14Drafter:
Luo Shaotang, Eugene HanDrafting Organization:
State-owned No. 721 FactoryFocal point Organization:
National Technical Committee for Standardization of Marine VesselsProposing Organization:
National Technical Committee for Standardization of Ferroelectric and Piezoelectric CeramicsPublishing Department:
General Administration of Quality Supervision, Inspection and Quarantine of the People's Republic of ChinaCompetent Authority:
China State Shipbuilding Corporation

Skip to download
Summary:
This standard specifies the test method for the Curie temperature Tc of piezoelectric ceramic materials. This standard is applicable to the test of the Curie temperature Tc of piezoelectric ceramic materials. GB/T 3389.3-2001 Test method for performance of piezoelectric ceramic materials Test of Curie temperature Tc GB/T3389.3-2001 Standard download decompression password: www.bzxz.net

Some standard content:
GB/T3389.3—2001
This standard is based on GB/T3389.3—1982 "Test Methods for Performance of Piezoelectric Ceramic Materials".
Compared with GB/T3389.3—1982, this standard has been revised as follows: 1 This standard adopts the mutation point of the test sample's permittivity to determine the Curie temperature Tc of the piezoelectric ceramic material. At the same time, it is recommended to use an impedance analyzer to test this mutation point;
2 Delete Appendix A Phase Transformer Measurement Method and Appendix B Transmission Line Measurement Method 3 Delete the provision that the two main planes of the sample are all covered with a metal layer as electrodes, and add the recommended sample size: @20mm×1mm.
This standard replaces GB/T3389.3—1982 from the date of implementation. This standard is proposed by the National Technical Committee for Standardization of Ferroelectric Piezoelectric Ceramics. This standard is under the jurisdiction of the National Technical Committee for Standardization of Ferroelectric Piezoelectric Ceramics. Drafting unit of this standard: State-owned No. 721 Factory. The main drafters of this standard are Luo Shaotang and Eugene Han. I
1 Scope
National Standard of the People's Republic of China
Test methods for the properties of piezoelectric ceramics-Test for Curie temperature Tc
Test methods for the properties of piezoelectric ceramics-Test for Curie temperature Te This standard specifies the test method for the Curie temperature Tc of piezoelectric ceramic materials. This standard is applicable to the test of the Curie temperature Tc of piezoelectric ceramic materials. 2 Referenced standards
GB/T 3389.3—2001
Replaces GB/T3389.3—1982
The provisions contained in the following standards constitute the provisions of this standard through reference in this standard. When this standard is published, the versions shown are valid. All standards will be revised, and parties using this standard should explore the possibility of using the latest versions of the following standards. GB/T3389.1—1996 Vocabulary of Ferroelectric and Piezoelectric Ceramics 3 Definitions and Symbols
The definitions and symbols used in this standard shall be in accordance with the provisions of GB/T3389.1. 4 Test Principle
The Curie temperature is a parameter that characterizes the ferroelectric properties of piezoelectric ceramic materials. Piezoelectric ceramics have different phase structures in different temperature ranges. The temperature at which phase transition occurs is defined as the phase transition temperature, and the highest phase transition temperature is called the Curie temperature. Above the Curie temperature, piezoelectric ceramics are in the paraelectric phase, and spontaneous polarization and piezoelectricity disappear. At the Curie temperature, many physical properties of piezoelectric ceramic materials, such as permittivity, heat capacity, and linear expansion coefficient, will undergo sudden changes. Therefore, the Curie temperature of piezoelectric ceramic materials can be determined by measuring the temperature corresponding to this mutation point. This standard uses the test specimen permittivity mutation point to determine the Curie temperature Tc of piezoelectric ceramic materials, and recommends the use of an impedance analyzer to measure this mutation point.
5 Specimen Requirements
The specimen is an unpolarized sheet with electrodes. The specimen should be kept clean and dry. Recommended sample size: Φ20mm×1mm.
6 Test method
6.1 Test circuit
The schematic diagram of the test circuit is shown in Figure 1.
Approved by the General Administration of Quality Supervision, Inspection and Quarantine of the People’s Republic of China on November 16, 2001 and implemented on August 1, 2002
6.2 Test equipment and requirements
Recorder
Impedance analyzer
Thermocouplewww.bzxz.net
6.3 Test procedure
GB/T 3389.3—2001
1-Recorder 2-Impedance analyzer 3-Electric furnace 4-Test sample 5-Thermocouple Figure 1 Block diagram of Curie temperature test circuit
The measurement error is not more than 1%;
The frequency variation is not more than ±0.5%, the output signal fluctuation is not more than ±1%, and the temperature in the furnace is uniform and controllable;
The measurement error is not more than 1%.
a) Place the sample in the furnace. The position of the thermocouple in the electric furnace should be able to truly reflect the temperature of the sample; b) Connect the circuit according to the test circuit shown in Figure 1, adjust the frequency of the impedance analyzer to a certain value, and then maintain its frequency unchanged; the electric furnace slowly heats up (or cools down). Under normal circumstances, the heating (or cooling) rate of the electric furnace is 3℃/min~4℃/min, and the rate near the Curie temperature is 1℃/min~2℃/min;
c) Read the peak value of the capacitance or susceptance versus temperature curve on the recorder. The temperature corresponding to this peak value is the Curie temperature of the material. If there is more than one peak on the curve, the corresponding temperatures are all phase change temperatures, and the highest phase change temperature is the Curie temperature. 2
Tip: This standard content only shows part of the intercepted content of the complete standard. If you need the complete standard, please go to the top to download the complete standard document for free.
This standard is based on GB/T3389.3—1982 "Test Methods for Performance of Piezoelectric Ceramic Materials".
Compared with GB/T3389.3—1982, this standard has been revised as follows: 1 This standard adopts the mutation point of the test sample's permittivity to determine the Curie temperature Tc of the piezoelectric ceramic material. At the same time, it is recommended to use an impedance analyzer to test this mutation point;
2 Delete Appendix A Phase Transformer Measurement Method and Appendix B Transmission Line Measurement Method 3 Delete the provision that the two main planes of the sample are all covered with a metal layer as electrodes, and add the recommended sample size: @20mm×1mm.
This standard replaces GB/T3389.3—1982 from the date of implementation. This standard is proposed by the National Technical Committee for Standardization of Ferroelectric Piezoelectric Ceramics. This standard is under the jurisdiction of the National Technical Committee for Standardization of Ferroelectric Piezoelectric Ceramics. Drafting unit of this standard: State-owned No. 721 Factory. The main drafters of this standard are Luo Shaotang and Eugene Han. I
1 Scope
National Standard of the People's Republic of China
Test methods for the properties of piezoelectric ceramics-Test for Curie temperature Tc
Test methods for the properties of piezoelectric ceramics-Test for Curie temperature Te This standard specifies the test method for the Curie temperature Tc of piezoelectric ceramic materials. This standard is applicable to the test of the Curie temperature Tc of piezoelectric ceramic materials. 2 Referenced standards
GB/T 3389.3—2001
Replaces GB/T3389.3—1982
The provisions contained in the following standards constitute the provisions of this standard through reference in this standard. When this standard is published, the versions shown are valid. All standards will be revised, and parties using this standard should explore the possibility of using the latest versions of the following standards. GB/T3389.1—1996 Vocabulary of Ferroelectric and Piezoelectric Ceramics 3 Definitions and Symbols
The definitions and symbols used in this standard shall be in accordance with the provisions of GB/T3389.1. 4 Test Principle
The Curie temperature is a parameter that characterizes the ferroelectric properties of piezoelectric ceramic materials. Piezoelectric ceramics have different phase structures in different temperature ranges. The temperature at which phase transition occurs is defined as the phase transition temperature, and the highest phase transition temperature is called the Curie temperature. Above the Curie temperature, piezoelectric ceramics are in the paraelectric phase, and spontaneous polarization and piezoelectricity disappear. At the Curie temperature, many physical properties of piezoelectric ceramic materials, such as permittivity, heat capacity, and linear expansion coefficient, will undergo sudden changes. Therefore, the Curie temperature of piezoelectric ceramic materials can be determined by measuring the temperature corresponding to this mutation point. This standard uses the test specimen permittivity mutation point to determine the Curie temperature Tc of piezoelectric ceramic materials, and recommends the use of an impedance analyzer to measure this mutation point.
5 Specimen Requirements
The specimen is an unpolarized sheet with electrodes. The specimen should be kept clean and dry. Recommended sample size: Φ20mm×1mm.
6 Test method
6.1 Test circuit
The schematic diagram of the test circuit is shown in Figure 1.
Approved by the General Administration of Quality Supervision, Inspection and Quarantine of the People’s Republic of China on November 16, 2001 and implemented on August 1, 2002
6.2 Test equipment and requirements
Recorder
Impedance analyzer
Thermocouplewww.bzxz.net
6.3 Test procedure
GB/T 3389.3—2001
1-Recorder 2-Impedance analyzer 3-Electric furnace 4-Test sample 5-Thermocouple Figure 1 Block diagram of Curie temperature test circuit
The measurement error is not more than 1%;
The frequency variation is not more than ±0.5%, the output signal fluctuation is not more than ±1%, and the temperature in the furnace is uniform and controllable;
The measurement error is not more than 1%.
a) Place the sample in the furnace. The position of the thermocouple in the electric furnace should be able to truly reflect the temperature of the sample; b) Connect the circuit according to the test circuit shown in Figure 1, adjust the frequency of the impedance analyzer to a certain value, and then maintain its frequency unchanged; the electric furnace slowly heats up (or cools down). Under normal circumstances, the heating (or cooling) rate of the electric furnace is 3℃/min~4℃/min, and the rate near the Curie temperature is 1℃/min~2℃/min;
c) Read the peak value of the capacitance or susceptance versus temperature curve on the recorder. The temperature corresponding to this peak value is the Curie temperature of the material. If there is more than one peak on the curve, the corresponding temperatures are all phase change temperatures, and the highest phase change temperature is the Curie temperature. 2
Tip: This standard content only shows part of the intercepted content of the complete standard. If you need the complete standard, please go to the top to download the complete standard document for free.
- Recommended standards
- GB/T 11711-2002 Performance requirements, test methods and required test results for marine automatic radar plotting aids (ARPA)
- GB 16334-1996 Practical dosimetry guide for food processing using gamma irradiation equipment
- QB/T 1683-1993 Aluminum-silicon-magnesium sagger
- YY/T 0457.6-2003 Characteristics of optoelectronic X-ray image intensifiers for medical electrical equipment Part 6: Determination of contrast and glare coefficient
- GB 325-1991 Packing container steel drum
- QB/T 2407-1998 Determination of D-panthenol content in cosmetics
- GB 2761-1981 Tolerance of Aflatoxin B1 in Food
- JJG 241-1981 Verification Regulation of Liquid Manometer (Grade Ⅱ&Ⅲ)
- SJ 20622-1997 General specification for military user entry node switches
- GB/T 3903.1-1994 General test methods for footwear - Folding test method
- GB 14886-1994 Road traffic signal installation specifications
- JB/T 8772.7-1998 Inspection conditions for precision machining centers Part 7: Accuracy inspection of precision machined test pieces
- GB 12487-2004 食品添加剂 乙基麦芽酚
- JB/T 3411.97-1999 Dimensions of end milling cutter bars for boring machines
- QB/T 2212.2-1996 General technical requirements for construction tools (brick and mortar)
Please remember: "bzxz.net" is the combination of the first letters of the Chinese pinyin of the four Chinese characters "standard download" and the international top-level domain name ".net". ©2024 Standard download websitewww.bzxz.net Mail:[email protected]