GB/T 11606.14-1989 Environmental test methods for analytical instruments - Low temperature storage test
time:
2024-08-06 10:58:15
- GB/T 11606.14-1989
- Abolished
Standard ID:
GB/T 11606.14-1989
Standard Name:
Environmental test methods for analytical instruments - Low temperature storage test
Chinese Name:
分析仪器环境试验方法 低温贮存试验
Standard category:
National Standard (GB)
-
Date of Release:
1989-03-31 -
Date of Implementation:
1990-03-01 -
Date of Expiration:
2007-12-01
Standard ICS number:
Test >> 19.040 Environmental TestChina Standard Classification Number:
Instruments and meters>>Material composition analysis instruments and environmental monitoring instruments>>N50 Material composition analysis instruments and environmental monitoring instruments comprehensive
alternative situation:
Replaced by GB/T 11606-2007
Review date:
2004-10-14Drafting Organization:
Beijing Analytical Instruments InstituteFocal point Organization:
National Technical Committee for Industrial Process Measurement and Control StandardizationPublishing Department:
China Machinery Industry FederationCompetent Authority:
China Machinery Industry Federation
Skip to download
Summary:
GB/T 11606.14-1989 Environmental test methods for analytical instruments Low temperature storage test GB/T11606.14-1989 Standard download decompression password: www.bzxz.net
Some standard content:
National Standard of the People's Republic of China
Environmental test method for analytical instruments
Cold storage test
The method of cnvironmental test for analytical instrumentscntsCold storage test
1 Subject content and scope of application
GB11606.14-89
This standard specifies the grouping, test conditions and test methods of low temperature storage test for analytical instruments (hereinafter referred to as instruments). This standard is applicable to the assessment of the adaptability of fully packaged instruments for transportation and storage under low temperature environmental conditions. 2 Low temperature storage test groupingwww.bzxz.net
3. Test conditions
3.1 Requirements for test equipment
3.1.1 The test box (room) should be able to provide the temperature specified in Chapter 2 of this standard in its working space, with an allowable error of ±2°C. The instrument should not be directly radiated by any heating and cooling elements. 3.1.2 The volume of the test box (room) should be greater than 3 times the volume of the instrument. 4 Test method
4.1 Pretreatment
Put the instrument in normal (or reference) environmental conditions to achieve temperature equilibrium. 4.2 Test
Put the instrument in the test box (room), then lower the temperature of the test box (room) to the specified value at an average temperature change rate of less than 1℃/min, and keep it for 8h.
4.3 Recovery
After the test, the instrument remains in the test box (room), and the temperature of the test box (room) is raised to normal (or reference) environmental conditions at an average temperature change rate of less than 1℃/min, and the recovery time is 8h. 4.4 Final inspection
Inspect the instrument according to relevant standards.
GB/T11606.14—1989 "Analytical Instrument Environmental Test Method Low Temperature Storage Test" Amendment No. 1 This amendment was approved by the State Administration of Quality and Technical Supervision on July 19, 2000 in the document No. [2000]108, and will be effective from December 1, 2000. 1. Add one more article between the original Articles 4.1 and 4.2, and it is numbered as Article 4.2. 4.2 Initial Inspection
Inspect the instrument according to the relevant standards. 2. The original Article 4.2 and the following articles are modified accordingly.
Tip: This standard content only shows part of the intercepted content of the complete standard. If you need the complete standard, please go to the top to download the complete standard document for free.
Environmental test method for analytical instruments
Cold storage test
The method of cnvironmental test for analytical instrumentscntsCold storage test
1 Subject content and scope of application
GB11606.14-89
This standard specifies the grouping, test conditions and test methods of low temperature storage test for analytical instruments (hereinafter referred to as instruments). This standard is applicable to the assessment of the adaptability of fully packaged instruments for transportation and storage under low temperature environmental conditions. 2 Low temperature storage test groupingwww.bzxz.net
3. Test conditions
3.1 Requirements for test equipment
3.1.1 The test box (room) should be able to provide the temperature specified in Chapter 2 of this standard in its working space, with an allowable error of ±2°C. The instrument should not be directly radiated by any heating and cooling elements. 3.1.2 The volume of the test box (room) should be greater than 3 times the volume of the instrument. 4 Test method
4.1 Pretreatment
Put the instrument in normal (or reference) environmental conditions to achieve temperature equilibrium. 4.2 Test
Put the instrument in the test box (room), then lower the temperature of the test box (room) to the specified value at an average temperature change rate of less than 1℃/min, and keep it for 8h.
4.3 Recovery
After the test, the instrument remains in the test box (room), and the temperature of the test box (room) is raised to normal (or reference) environmental conditions at an average temperature change rate of less than 1℃/min, and the recovery time is 8h. 4.4 Final inspection
Inspect the instrument according to relevant standards.
GB/T11606.14—1989 "Analytical Instrument Environmental Test Method Low Temperature Storage Test" Amendment No. 1 This amendment was approved by the State Administration of Quality and Technical Supervision on July 19, 2000 in the document No. [2000]108, and will be effective from December 1, 2000. 1. Add one more article between the original Articles 4.1 and 4.2, and it is numbered as Article 4.2. 4.2 Initial Inspection
Inspect the instrument according to the relevant standards. 2. The original Article 4.2 and the following articles are modified accordingly.
Tip: This standard content only shows part of the intercepted content of the complete standard. If you need the complete standard, please go to the top to download the complete standard document for free.
- Recommended standards
- JB/T 9897-1999 Accuracy test of universal radial drilling machine without base
- GB/T 7706-2008 The relief prints for decorating
- JB/T 6720-1993 Metallographic inspection of intake and exhaust valves of internal combustion engines
- JB/T 10127.2-1999 Machine tool fixture parts and components - Isosceles angle iron
- GB 4578-1984 Food additive Saccharin sodium
- GB/T 8508-1987 Reliability test method for tire rollers
- GB/T 15072.5-1994 Chemical analysis methods for precious metals and their alloys - Determination of silver content in gold and palladium alloys
- SL 77-1994 Specification for hydrological calculations for small hydroelectric power stations SL 77-94
- HG/T 2277-1992 Various colors of nitrocellulose external enamel
- JB/T 2231.1-1999 Reciprocating piston compressor parts Part 1: Shaft and pin outer diameter dimensions
- JB/T 6443-2002 Centrifugal compressors for petroleum, chemical and gas industries
- GB/T 19629-2005 Medical electrical equipment - Ionization chambers and/or semiconductor detector dosimeters used in X-ray diagnostic imaging
- RB/T 241-2018 Requirements for laboratory in the field of green products testing
- HG/T 2048.2-1991 Stuffing box with cooling water jacket
- HG 3322-1981 Determination of permanent deformation of vulcanized rubber (modulus tester method)
Please remember: "bzxz.net" is the combination of the first letters of the Chinese pinyin of the four Chinese characters "standard download" and the international top-level domain name ".net". ©2024 Standard download websitewww.bzxz.net Mail:bzxznet@163.com