JB/T 5528-1991 Pressure gauge scale and divisions
time:
2024-06-24 10:26:52
- JB/T 5528-1991
- Abolished
Standard ID:
JB/T 5528-1991
Standard Name:
Pressure gauge scale and divisions
Chinese Name:
压力表 标度及分划
Standard category:
Machinery Industry Standard (JB)
-
Date of Implementation:
1992-07-01 -
Date of Expiration:
2006-02-01
China Standard Classification Number:
Instruments and meters>>Industrial automation instruments and control devices>>N11 temperature and pressure instrument
alternative situation:
Replaced by JB/T 5528-2005

Skip to download
Summary:
JB/T 5528-1991 Pressure gauge scale and division JB/T5528-1991 standard download decompression password: www.bzxz.net

Some standard content:
Mechanical Industry Standard of the People's Republic of China
JB/T5528--91
Pressure Gauge Scale and Division
1991-07-16 Issued
Ministry of Machinery and Electronics Industry
1992-07-01 Implementation
Subject Content and Scope of Application
Mechanical Industry Standard of the People's Republic of China
Pressure Gauge Scale and Division
This standard specifies the dial scale and division of various pressure gauges. JB/T5528-91
This standard is applicable to the scale design of the dial of various pressure gauges (referred to as instruments) with accuracy levels of 0.1, 0.16, 0.25, 0.4, 1, 1.5, 2.5, 4 and shell nominal diameters of Φ40, 60, $100, Φ150, 200, Φ250mm. Pressure gauges with other types of shells can also be used as a reference.
Those not specified in this standard shall comply with the corresponding provisions of ZBY123. 2
Cited standards
ZBY123 General provisions for the scale of industrial automation instruments 3 Basic requirements for scale
3.1 Scale direction
The scale direction of the instrument shall comply with the following provisions; a. The zero scale line of the pressure gauge shall be located at the left end or lower end of the scale, and scale from zero to the right, clockwise or upward. b. The zero scale line of the vacuum gauge shall be located at the right end or upper end of the scale, and scale from zero to the left, counterclockwise or downward. C. The zero scale line of the pressure vacuum gauge shall be located between the vacuum part scale and the pressure part scale, and the vacuum part scale shall be scaled from zero to the left, counterclockwise or downward, and the pressure part scale shall be scaled from zero to the right, clockwise or upward. 3.2 Zero scale line
The scale reduction of the actual zero scale line of the instrument dial with a stop pin shall not be greater than the absolute value of the basic error limit of the indication. The zero scale line of the instrument dial without a stop pin shall be indicated along the scale baseline, and the indicated range shall not be greater than twice the absolute value of the basic error limit of the indication. 3.3 Scale line
3.3.1 For instruments with accuracy levels of 1, 1.5, 2.5, and 4, the width of the long scale line, the middle scale line, and the short scale line shall not be greater than that specified in Table 1. The long scale line uses a semi-thick line, and the width of the thick part of the semi-thick line shall not be greater than 3 times that specified in Table 1. Table 1
Nominal shell diameter
Approved by the Ministry of Machinery and Electronics Industry on July 16, 1991 Width of scale lines
Implemented on July 1, 1992
JB/T5528-91
3.3.2 For instruments with an accuracy level of 0.4 or higher, the scale lines shall be all fine lines with a width not exceeding 0.2mm. 3.4 Scale units
According to the measuring range of the instrument, the scale units may be MPa (megapascals), kPa (kilopascals) or Pa (pascals). The specific selection of scale units shall be specified by the product standards.
Scale divisions
4.1 The scale diagrams of instruments with a nominal shell diameter of Φ40, (50), @60mm and an accuracy level of 2.5 and 4 shall be as specified in Table 2. Note: The shell diameter in brackets can only be used for pressure gauges used for welding, cutting and similar processes. 4.2 The nominal shell diameter is Φ100mm, and the instrument scales of the accuracy levels 1, 1.5, and 2.5 are as specified in Table 3. 4.3 The nominal shell diameter is 150, Φ200, and Φ250mm, and the instrument scales of the accuracy levels 1, 1.5, and 2.5 are as specified in Table 4. 4.4 The nominal shell diameter is Φ150, 200, and Φ250mm, and the instrument scales of the accuracy level 0.4 are as specified in Table 5. The nominal shell diameter is Φ150, Φ200, and Φ250mm, and the instrument scales of the accuracy level 0.25 are as specified in Table 6. 4.5
The nominal shell diameter is Φ250mn, and the instrument scales of the accuracy levels 0.1 and 0.16 are as specified in Appendix A (reference). Table 2
Nominal shell diameter Φ40 (50) 60mm, accuracy level 2.5, 4 instrument scale schematic scale range
—10~0
-0.05~0.05
Minimum grid value
Total number of grids
Scale range
0:16~0
Minimum grid value
Total number of grids
JB/T5528--91||tt ||Continued Table 2
Scale range
0.12~0.12
—1.2~1.2
~12~12
20~20
Minimum grid value
Total number of grids
5528-91
Continued Table 2
Scale range
-6~0
60~0
-3~3
-30~30
Scale range||t t||Minimum grid value
Total number of grids
JB/T5528-91
Continued Table 2
Scale diagram of instruments with nominal shell diameter Φ100mm and accuracy levels 1, 1.5, and 2.5Minimum grid value
Total number of grids
Scale range
0.03~0.03
-10~0
0.05~0.05
-5~5
Minimum grid value||t t||Total number of grids
JB/T5528-91
Continued Table 3
Scale range
-16~0
0.08~0.08
-25~0
Minimum grid value
Total number of grids
JB/T5528—91
Continued Table 3
Scale range
-0.12~0.12
-12~~12
-4~ 0
-0.2~~0.2
-20~20
Minimum grid value
Total number of grids
JB/T5528--91
Continued Table 3
JB/T5528—91
Nominal shell diameter 150, 9
200, Φ250mm,
Scale schematic scale range of instruments with accuracy levels 1, 1.5, and 2.5
—0.06~0
-6~0
-0.03~0.03| |tt||-0.3~0.3
-3~3
-30~30
-1~0
Minimum grid value
Total number of grids
titjtiWww.bzxZ.net
Scale range
0.05~0.05
-5~5
16~0
0.08~0.08
-0.1~0.06
Minimum grid value
Total number of grids
JB/T5528--91
Continued Table 4
Tip: This standard content only shows part of the intercepted content of the complete standard. If you need the complete standard, please go to the top to download the complete standard document for free.
JB/T5528--91
Pressure Gauge Scale and Division
1991-07-16 Issued
Ministry of Machinery and Electronics Industry
1992-07-01 Implementation
Subject Content and Scope of Application
Mechanical Industry Standard of the People's Republic of China
Pressure Gauge Scale and Division
This standard specifies the dial scale and division of various pressure gauges. JB/T5528-91
This standard is applicable to the scale design of the dial of various pressure gauges (referred to as instruments) with accuracy levels of 0.1, 0.16, 0.25, 0.4, 1, 1.5, 2.5, 4 and shell nominal diameters of Φ40, 60, $100, Φ150, 200, Φ250mm. Pressure gauges with other types of shells can also be used as a reference.
Those not specified in this standard shall comply with the corresponding provisions of ZBY123. 2
Cited standards
ZBY123 General provisions for the scale of industrial automation instruments 3 Basic requirements for scale
3.1 Scale direction
The scale direction of the instrument shall comply with the following provisions; a. The zero scale line of the pressure gauge shall be located at the left end or lower end of the scale, and scale from zero to the right, clockwise or upward. b. The zero scale line of the vacuum gauge shall be located at the right end or upper end of the scale, and scale from zero to the left, counterclockwise or downward. C. The zero scale line of the pressure vacuum gauge shall be located between the vacuum part scale and the pressure part scale, and the vacuum part scale shall be scaled from zero to the left, counterclockwise or downward, and the pressure part scale shall be scaled from zero to the right, clockwise or upward. 3.2 Zero scale line
The scale reduction of the actual zero scale line of the instrument dial with a stop pin shall not be greater than the absolute value of the basic error limit of the indication. The zero scale line of the instrument dial without a stop pin shall be indicated along the scale baseline, and the indicated range shall not be greater than twice the absolute value of the basic error limit of the indication. 3.3 Scale line
3.3.1 For instruments with accuracy levels of 1, 1.5, 2.5, and 4, the width of the long scale line, the middle scale line, and the short scale line shall not be greater than that specified in Table 1. The long scale line uses a semi-thick line, and the width of the thick part of the semi-thick line shall not be greater than 3 times that specified in Table 1. Table 1
Nominal shell diameter
Approved by the Ministry of Machinery and Electronics Industry on July 16, 1991 Width of scale lines
Implemented on July 1, 1992
JB/T5528-91
3.3.2 For instruments with an accuracy level of 0.4 or higher, the scale lines shall be all fine lines with a width not exceeding 0.2mm. 3.4 Scale units
According to the measuring range of the instrument, the scale units may be MPa (megapascals), kPa (kilopascals) or Pa (pascals). The specific selection of scale units shall be specified by the product standards.
Scale divisions
4.1 The scale diagrams of instruments with a nominal shell diameter of Φ40, (50), @60mm and an accuracy level of 2.5 and 4 shall be as specified in Table 2. Note: The shell diameter in brackets can only be used for pressure gauges used for welding, cutting and similar processes. 4.2 The nominal shell diameter is Φ100mm, and the instrument scales of the accuracy levels 1, 1.5, and 2.5 are as specified in Table 3. 4.3 The nominal shell diameter is 150, Φ200, and Φ250mm, and the instrument scales of the accuracy levels 1, 1.5, and 2.5 are as specified in Table 4. 4.4 The nominal shell diameter is Φ150, 200, and Φ250mm, and the instrument scales of the accuracy level 0.4 are as specified in Table 5. The nominal shell diameter is Φ150, Φ200, and Φ250mm, and the instrument scales of the accuracy level 0.25 are as specified in Table 6. 4.5
The nominal shell diameter is Φ250mn, and the instrument scales of the accuracy levels 0.1 and 0.16 are as specified in Appendix A (reference). Table 2
Nominal shell diameter Φ40 (50) 60mm, accuracy level 2.5, 4 instrument scale schematic scale range
—10~0
-0.05~0.05
Minimum grid value
Total number of grids
Scale range
0:16~0
Minimum grid value
Total number of grids
JB/T5528--91||tt ||Continued Table 2
Scale range
0.12~0.12
—1.2~1.2
~12~12
20~20
Minimum grid value
Total number of grids
5528-91
Continued Table 2
Scale range
-6~0
60~0
-3~3
-30~30
Scale range||t t||Minimum grid value
Total number of grids
JB/T5528-91
Continued Table 2
Scale diagram of instruments with nominal shell diameter Φ100mm and accuracy levels 1, 1.5, and 2.5Minimum grid value
Total number of grids
Scale range
0.03~0.03
-10~0
0.05~0.05
-5~5
Minimum grid value||t t||Total number of grids
JB/T5528-91
Continued Table 3
Scale range
-16~0
0.08~0.08
-25~0
Minimum grid value
Total number of grids
JB/T5528—91
Continued Table 3
Scale range
-0.12~0.12
-12~~12
-4~ 0
-0.2~~0.2
-20~20
Minimum grid value
Total number of grids
JB/T5528--91
Continued Table 3
JB/T5528—91
Nominal shell diameter 150, 9
200, Φ250mm,
Scale schematic scale range of instruments with accuracy levels 1, 1.5, and 2.5
—0.06~0
-6~0
-0.03~0.03| |tt||-0.3~0.3
-3~3
-30~30
-1~0
Minimum grid value
Total number of grids
titjtiWww.bzxZ.net
Scale range
0.05~0.05
-5~5
16~0
0.08~0.08
-0.1~0.06
Minimum grid value
Total number of grids
JB/T5528--91
Continued Table 4
Tip: This standard content only shows part of the intercepted content of the complete standard. If you need the complete standard, please go to the top to download the complete standard document for free.
- Recommended standards
- GB/T 5211.1-2003 Determination of water-soluble matter in pigments - Cold extraction method
- JB/T 7549.3-1994 Roller sealing machine
- QB/T 2250-2005 Single-sided white paperboard
- HG/T 3139.1-2001 Types and basic parameters of vertical reducers for kettles
- GB/T 4501-1998 Truck tire durability test method - Drum method
- JB/T 9212-1999 Ultrasonic flaw detection of welds of atmospheric pressure steel oil tanks
- SY/T 5023-1994 Brake pads for oil drilling rigs
- GB/T 5474-1985 Determination of free ammonia and ammonium compounds in phenolic mouldings - Colorimetric method
- JB/T 1373-1974 Technical requirements for electrical contacts
- JB/T 8771.1-1998 Inspection conditions for machining centres Part 1: Geometric accuracy inspection of horizontal and additional spindle head machines (horizontal Z axis)
- QB/T 3507-1999 Computer continuous recording format base paper
- JB/T 2104-2002 Hydraulic jack
- GB 11920-1989 General technical conditions for centralized control devices of electrical parts of power stations
- GB/T 38568-2020 Determination of growth phenotype for industrialmicrobial strain—Microdroplet turbidity
- GB/T 16716.2-2010 Packaging and packaging waste—Part 2:Means and program for evaluation
Please remember: "bzxz.net" is the combination of the first letters of the Chinese pinyin of the four Chinese characters "standard download" and the international top-level domain name ".net". ©2024 Standard download websitewww.bzxz.net Mail:bzxznet@163.com